The most substantial novelty certainly concerns the definition of the time intervals within which the glove test must be performed, thus defining an aspect that until now was at the discretion of the manufacturer. In fact, the first version of Annex 1, drafted in 2007, states that glove leak tests must take place frequently, but without giving any further specifications.
The newly published revision fills precisely this gap, specifying instead that, as far as isolators are concerned, glove testing must be performed at defined intervals, at least at the beginning and end of each production batch.
"The testing should be performed at defined intervals. Generally, glove integrity testing should be performed at a minimum frequency of the beginning and end of each batch or campaign.1"In the case of very small hand-processed productions, the criterion is no longer the batch, but rather the manufacturing session.
“For manual aseptic processing activities where single unit or small batch sizes are produced, the frequency of integrity verification may be based on other criteria, such as the beginning and end of each manufacturing session.1”Clearly, such a change in the regulatory framework cannot fail to have an impact on manufacturing processes. As of 25 August 2023, when the new Annex 1 will come into force, pharmaceutical companies within the European territory, or intending to market their products in this market, will in fact be faced with the need to increase the frequency of glove tests. It is inevitable that in such a scenario, the method used and the technologies adopted to perform the test will play a key role.
As we have seen, the choice of the right glove integrity testing equipment is a key aspect within the manufacturing process. As it 'demonstrated to be suitable for the task and criticality', as the revised Annex 1 states, there are also some technical and usage characteristics that should not be underestimated, with reference to the new regulatory framework.
Let's see some of them.
A final aspect not to be underestimated is the ergonomics and user-friendliness of the equipment. Certainly a user-friendly instrument will enable the operator to perform the required operations at best in the shortest possible time. But this is not the only aspect to be taken into consideration. Maintenance is also crucial. Specifically cleaning. Since this is an instrument used almost exclusively in classified environments, it is advisable to evaluate this aspect as well, since an easy-to-clean instrument will take less time to be ready for use, thus saving time and resources.
| 1 Quote from GUIDELINES | The Rules Governing Medicinal Products in the European Union Volume 4 EU Guidelines for Good Manufacturing Practice for Medicinal Products for Human and Veterinary Use Annex1 |